Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/07/1998 | 1430 | 05/07/1998 | 1500 | Incorrect |
Subject: | SGP/EBBR/E19 - Six-Month Checks |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Suggestions: | |
Measurements: | sgp30ebbrE19.b1: more
sgp15ebbrE19.b1: more sgp5ebbrE19.b1: more |