Subject: | SGP/QCRAD/E15 - Temp and RH measurements incorrect |
DataStreams: | sgpqcrad1longE15.c1, sgpqcrad1longE15.c2
|
Description: | Surface meteorological data were missing during these time periods. Met data is used in a
variety of quality control checks described below.
Air temperature data is needed for tests on LW up and LW down pyrgeometer case and dome
temperatures, and tests on the irradiances from upwelling LW/SW and downwelling LW.
Most periods with missing data described in instrument DQRs: D980521.1, D980616.1 |
Suggestions: | Use caution with LW up and down pyrgeometer irradiance, dome temperature, case
temperature, SW up data and diffuse irradiance data because all quality control checks could
not be run. |
Measurements: | sgpqcrad1longE15.c2: - LWupTc
- rh
- LWdnTc
- LWdnTd
- LWupTd
- down_short_diffuse_hemisp
- Temp_Air
- down_long_hemisp
- up_short_hemisp
- up_long_hemisp
more
sgpqcrad1longE15.c1: - up_long_hemisp
- down_long_hemisp
- LWdnTd
- up_short_hemisp
- rh
- Temp_Air
- LWupTd
- LWupTc
- LWdnTc
- down_short_diffuse_hemisp
more
|