Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/08/1998 | 2051 | 12/15/1998 | 1200 | Suspect |
Subject: | NSA/GNDRAD/C1 - PIR1 Calibration problems resulted in small error |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The dome correction coefficients were not properly reset in calibration file until the early morning of 12/15. These errors probably account for less than 4% in error. |
Suggestions: | |
Measurements: | nsagndrad60sC1.a1: more
|