Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/23/1996 | 1610 | 04/23/1996 | 1703 | Incorrect |
Subject: | SGP/EBBR/E15 - 3-month check |
DataStreams: | sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Suggestions: | |
Measurements: | sgp5ebbrE15.b1: more
sgp15ebbrE15.b1: more sgp30ebbrE15.b1: more |