Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/01/1993 | 0000 | 12/21/1993 | 1800 | Incorrect |
Subject: | SGP/EBBR/E22 - Air Temperature Sensor malfunction |
DataStreams: | sgp30ebbrE22.b1, sgp15ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The right air temperature probe (thermocouple) malfunctioned during the periods stated above at Cordell, OK (E22). It was replaced on December 21, 1993 and good data began again with the 1830 GMT half hour average. For the periods of interest, 5 and 30 minute values of tair_top and tair_bot are incorrect when the right probe was in the specified position; 15 minute values of tair_r are incorrect; and 30 minute values of bowen, e, and h are incorrect. None of these values are recoverable or recalculatable through the use of other quantities since both air temperature probes are needed to determine bowen ratio. |
Suggestions: | |
Measurements: | sgp15ebbrE22.b1: more
sgp5ebbrE22.b1: more sgp30ebbrE22.b1: more |